MODAPS PCR 11-041

PCR Number 11-041
Date 2011-09-07
Initiator James Kuyper
Abstract This Aqua LUT patch PGE02 5.0.41_1 (recipes PM1M_1(nrt)m(c) 3.3.99) includes a new algorithm to calculate the default b1 on-orbit considering the impact of the real time LWIR FPA temperature.
Problem Statement The current Aqua L1B code uses fixed b1 values for bands 33, 35 and 36 when the blackbody temperature is above a certain threshold. LWIR FPA temperature variations have recently increased due to the loss of radiative cooler margin. This is only a problem during warm-up/cool-down calibration tests, which last about 4 hours and happen only once every 3 months.
Description of Change An algorithm to calculate temperature-dependent b1 values was developed by MCST. The new default b1 is calculated as:b1 = default_b1_baseline*(1+c1*(Tlwir – Tlwir_baseline))Where default_b1_baseline, c1 and Tlwir_baseline are provided by MCST through 3 new time-dependent LUTs, Tlwir is the real time LWIR FPA temperature. One old lut bb_t_sat_default_b1_aqua_table is removed.
Products Affected MYD021KM (recipes PM1M_1(nrt)m(c) 3.3.99)
Software Affected PGE02 v5.0.41
Processing String to Receive the Change Aqua Forward
Downstream Product Effects Any downstream product using Aqua bands 33, 35 and 36 L1B data may be affected. The impact only occurs during blackbody warm up/cool down events, and should be about 1%.
Data Granules to be Used for Science Testing Aqua granules between 2011233.0840 and 2011233.1315.2010-059
Justification The internal tests requested by MCST have demonstrated that the new algorithm accurately adjusts for the impact of the LWIR FPA temperature changes.
Effective Date for Implementation of Change 2012-04-15
Status Bill Ridgway and Chris Moeller approved for Atmospheres, Robert Wolfe approved for Land, Michael King approved on 04/06/2012.